VHP BIBO Sterilizer
A vaporized hydrogen peroxide (VHP) sterilizer that decontaminates bag-in/bag-out (BIBO) HEPA filter housings in place, so contaminated filters can be changed safely.
Request a Quote All ProductsOverview
The Fenggu VHP BIBO Sterilizer performs in-situ vaporized hydrogen peroxide biodecontamination of bag-in/bag-out (BIBO) HEPA filter housings after normal filtration service, before a filter change, so that filters can be handled and replaced safely. The cycle runs in four automatic steps — dehumidification, conditioning, exposure and aeration. A molecular-sieve desiccant adsorbs moisture, hydrogen peroxide vapor is then introduced at a stated concentration of approximately 30 ppm for approximately 40 minutes, and residual vapor is removed by recirculation through a palladium catalyst, or by operating the BIBO supply and exhaust system, until the concentration is below 1 ppm. Cycles can be developed to a target of at least a 6-log reduction of Geobacillus stearothermophilus spores; actual efficacy is subject to the project validation protocol and report.
Key Features
- Four automatic steps: dehumidification / conditioning / exposure / aeration
- Molecular-sieve dehumidification to < 30% RH before gassing
- ≥6-log reduction of Geobacillus stearothermophilus spores targeted (approx. 30 ppm, ~40 min; subject to validation)
- Residual VHP removed via palladium catalyst recirculation or the BIBO supply/exhaust system
- Siemens PLC control with Siemens HMI / tablet touchscreen
- High-resolution monitoring: humidity 0.1% / temperature 0.1°C / pressure 0.1 Pa
Technical Specifications
| Overall dimensions | 640 × 836 × 1036 mm | Sporicidal efficacy | ≥6-log reduction (subject to validation) |
|---|---|---|---|
| Metal material | SUS304 / SUS316L | Non-metal material | PVC, silicone |
| Air flow | ≤ 500 L/min | Treatment volume | ≤ 30 m³ |
| Vaporization temperature | ≤ 120°C | Dehumidification | < 30% RH |
| Sterilant | 30% or 35% H2O2 | Residual VHP | ≤ 1 ppm (catalytic) |
| Power supply | AC 220 V, 50 Hz, ≥ 3 kW | Control | Siemens PLC + HMI / tablet |
| Noise | ≤ 75 dB(A) | Weight | approx. 80 kg |
Applications
Request a datasheet or quotation
Tell us the BIBO housing type and filter configuration, and our engineers will recommend a suitable VHP BIBO sterilizer setup.