[email protected]   +86 183 2145 9135EN|РУ|中文Inquiry Now
HomeProducts › Automatic-Door VHP Pass Box
Transfer & Containment FG-PB-VHPA

Automatic-Door VHP Pass Box

An automated VHP transfer chamber combining motorized door operation, interlocking and validated surface bio-decontamination.

Request a Quote All Products
Fenggu Automatic-Door VHP Pass Box FG-PB-VHPA
Fenggu Automatic-Door VHP Pass Box FG-PB-VHPA
Fenggu Automatic-Door VHP Pass Box FG-PB-VHPA
Fenggu Automatic-Door VHP Pass Box FG-PB-VHPA

Overview

The Fenggu Automatic-Door VHP Pass Box combines vaporized hydrogen peroxide surface bio-decontamination with automatic door control for material transfer between controlled environments. Motorized door operation reduces manual contact and supports repeatable transfer workflows in pharmaceutical, biosafety and healthcare facilities.

The integrated system monitors critical process conditions, records cycle data and reduces residual hydrogen peroxide through catalytic decomposition before door release. The two doors are interlocked so they cannot be opened simultaneously.

Key Features

  • Automatic door opening and closing with safety interlocking
  • VHP surface bio-decontamination with a documented 6-log reduction target
  • Inflatable or mechanical sealing configuration according to project requirements
  • Monitoring of temperature, humidity and hydrogen peroxide concentration
  • Catalytic decomposition to reduce residual hydrogen peroxide to ≤1 ppm
  • H14 HEPA-filtered inlet and outlet air paths
  • Siemens PLC automatic control and cycle data recording

Technical Parameters

External dimensionsConfigured to project requirementsChamber materialSUS304
Internal pipingSUS316LBio-decontamination efficacy6-log reduction
Process temperature25–35 °CProcess humidity40–60% RH
VHP concentration300–400 ppmPost-cycle residual≤1 ppm
Bio-decontamination cycle30 min (documented nominal value)Aeration time≤1 h
Door systemAutomatic, double-door interlockedAir filtrationH14 HEPA
Door type, chamber dimensions, load arrangement and cycle recipe are configured to the application. Final cycle performance must be confirmed during site-specific qualification and validation.

Applications

Sterile Material Transfer Pharmaceutical Production Biosafety Laboratories Healthcare Clean Areas High-Throughput Transfer

Request a datasheet or quotation

Tell us your chamber size, door arrangement, load type and cycle requirements, and our engineers will recommend a configuration.

Contact Our Team

Frequently Asked Questions

What is the advantage of an automatic-door VHP pass box?
Motorized double-door operation reduces manual contact and gives repeatable, hands-free transfer workflows, while an in-chamber VHP cycle surface-decontaminates materials between controlled environments.
What are the cycle parameters?
Process 25-35 C, 40-60% RH, VHP 300-400 ppm; a roughly 30-minute bio-decontamination cycle, aeration of 1 hour or less to a residual of 1 ppm or less; designed for a 6-log reduction.
Are the doors interlocked?
Yes. The automatic double doors are interlocked so both sides cannot open at once, maintaining the pressure and containment boundary. Air is H14 HEPA filtered.
Can it be customized?
Door type, chamber dimensions, load arrangement and cycle are configured to your materials and the grades of the connected rooms. Construction is SUS304 with SUS316L piping.